![Datex-Ohmeda S/5 Compact Anesthesia Monitor Technical Reference Manual Download Page 55](http://html.mh-extra.com/html/datex-ohmeda/s-5-compact-anesthesia-monitor/s-5-compact-anesthesia-monitor_technical-reference-manual_513839055.webp)
General Service Guide
1.
Block the tip of the sampling line with your finger and check that the message 'Sample line
blocked' appears on the monitor screen within 30 seconds.
2.
Detach the Mini-D-fend and check that the message 'Check D-fend' appears on the
monitor screen within 30 seconds.
3.
Breathe to the sampling line briefly. Check that the CO2 information is updated on the
screen.
"
5.3.12 Tonometry Module, M-TONO
1. Press each of the front panel membrane keys for at least one second and check that they are
identified.
"
5.3.13 Hemodynamic Modules
ECG and RESP measurements
1. Check functioning of the front panel membrane key ECG LEAD, if available.
2. Connect an ECG cable to the module. Connect the cable leads to a patient simulator. Check
that all ECG and impedance respiration information is shown on the monitor screen as
configured on the simulator.
Turn the simulator off. Check that the ‘Asystole ’ and ‘Apnea’ messages are displayed.
"
Modules w/temperature measurement
3. Check the temperature channels with a patient simulator.
Check that temperature measurement information is shown on the monitor screen as
configured on the simulator.
"
Modules w/invasive blood pressure measurement
4. Check functioning of the front panel membrane keys.
5. Check the InvBP channels with a patient simulator.
Zero the InvBP channels, the check that the values and waveforms correspond to the
simulator settings.
"
Modules w/SpO2 measurement
6. Connect SpO2 finger sensor to the module. Check that the message ‘Probe off ’ is show when
the sensor is not connected to a finger.
7. Attach the SpO2 sensor to your finger. Check that a reading of 95-99 and a Pleth waveform
appear onto the screen
"
Document No. M1031508
45
Summary of Contents for S/5 Compact Anesthesia Monitor
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 ii ...
Page 10: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 iv ...
Page 60: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 50 ...
Page 61: ...General Service Guide APPENDICES A B Document No M1031508 51 ...
Page 62: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 52 ...
Page 72: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 B 6 6 ...