General Service Guide
4.2.5 Selecting the external device
1.
Connect the interfacing module to the monitor’s DIS connector or to another interfacing
module.
2.
Connect the device specific cable to the external device and turn the external device on.
3.
The monitor identifies the connected device automatically.
4.2.6 Selecting the parameter data source
Select the external device via
Monitor Setup
-/
Interfacing
– menu:
•
Select the desired measurement parameter (e.g.,
Gases
).
•
Select the desired source by name (e.g.,
Aestiva
).
NOTE: The name of the device is visible on the list only if the device is correctly connected.
4.3 Interfacing Datex-Ohmeda S/5 Anesthesia Delivery Unit
It is possible to interface the Datex-Ohmeda S/5 Anesthesia Delivery Unit with the S/5 Compact
Anesthesia Monitor via the serial interface port. The data link is bi-directional. The parameters
transferred to the Anesthesia Delivery Unit are summarized in table 3 and the events transferred to
the S/5 Compact Anesthesia Monitor are summarized in table 4. The events are transferred
continuously once a minute or during the system check procedure and appear in the trend pages or
in the anesthesia record.
4.3.1 Interconnection
Use the ADU-Monitor Cable.
4.3.2 Setting interfacing parameters for the S/5 Anesthesia Delivery Unit
1. Press
the
Setup
key.
2. Select
Install
password 10.
3. Select
Interfacing
and choose S/5 AM.
4. Press
the
Normal Screen
key.
Document No. M1031508
31
Summary of Contents for S/5 Compact Anesthesia Monitor
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 ii ...
Page 10: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 iv ...
Page 60: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 50 ...
Page 61: ...General Service Guide APPENDICES A B Document No M1031508 51 ...
Page 62: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 52 ...
Page 72: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 B 6 6 ...