Datex-Ohmeda S/5 Compact Anesthesia and Critical Care Monitor
Check that the corresponding icon appears onto the monitor screen.
"
5.3.8 Network Option
1.
Check that the Mon-Net cable connector and the Identification plug are clean and intact,
then connect them to the frame unit.
Check that the monitor connects onto the S/5 Network, i.e. the network symbol appears on
the upper right-hand corner of the screen.
"
5.3.9 Wireless LAN Option
1.
Check that the green LED on the wireless antenna card blinks and the WLAN signal strength
symbol roams or stays fixed on the monitor screen.
2.
Check that the monitor connects onto the S/5 Network, i.e. the wireless LAN network
symbol appears on the upper right-hand corner of the screen.
"
NOTE: If the monitor does not connect onto the S/5 network, check the WLAN configuration on the
monitor and on the network.
5.3.10 Compact Airway Module, M-CXXXXX
Wait until the message ‘Calibrating gas sensor ’ disappears from the screen.
1. Check that the fan is running.
2. If the module contains membrane keys on the front panel, press each of the keys for at least
one second and check that they are identified.
"
Compact Airway Modules w/the Patient Spirometry option
3. Connect a clean spirometry tube and D-lite to the module. Connect the sampling line.
Breathe through the wider side of the D-lite. Check that the flow waveform moves downwards
when you breathe in and upwards when you breathe out.
"
For all Compact Airway Modules
4. Block the tip of the sampling line with your finger and check that the message ‘Sample line
blocked ’ appears on the monitor screen within 30 seconds.
5. Detach the D-fend and check that the messages ‘Check D-fend ’ appears on the monitor
screen within 30 seconds.
6. Breathe to the sampling line briefly. Check that the CO2 information is updated on the screen.
"
5.3.11 Single width Airway Module, M-miniC
Wait until the message 'Calibrating gas sensor' disappears from the screen.
Document No. M1031508
44
Summary of Contents for S/5 Compact Anesthesia Monitor
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 ii ...
Page 10: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 iv ...
Page 60: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 50 ...
Page 61: ...General Service Guide APPENDICES A B Document No M1031508 51 ...
Page 62: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 52 ...
Page 72: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 B 6 6 ...