4.1.3 Connection to Baxter Vigilance
Use the INT-Baxter Vigilance Cable.
1. Make sure that the power on both monitors is switched off.
2. Connect the 9 pin D-connector to one of the connectors on the Interface Module, M-INT.
Tighten the finger screws.
3. Connect the other 9 pin D-connector to the connector X3, X7, X8 or X9 on the interfaced
monitor. Tighten the finger screws.
Communication parameters: Baud rate 19200, no parity (none), data bits 8, stop bits 1.
Mode of communication port: IFM out.
4.1.4 Connection to Nellcor N-100 and N-1000
Use the Monitor-Nellcor Cable.
1. Make sure that the power on both monitors is switched off.
2. Connect the 9 pin D-connector to one of the connectors on the Interface Module, M-INT.
Tighten the finger screws.
3. Connect the mono connectors to the corresponding connectors on the interfaced monitor.
4.1.5 Connection to Nellcor N-200
Use the Monitor-Nellcor Cable.
1.
Make sure that the power on both monitors is switched off.
2.
Connect one of the connectors to the corresponding connector on the Interface Module, M-
INT. Tighten the finger screws.
3.
Connect the other connector to the corresponding connector on the interfaced monitor.
Tighten the finger screws.
4.
Connect the mono connectors to the corresponding connectors on the interfaced monitor.
4.1.6 Setting S/5 Compact Anesthesia Monitor and S/5 Compact Critical Care Monitor
interfacing parameters
1. Press
the
Monitor Setup
key.
2. Select
Install/Service
(password 16-4-34).
3. Select
Installation
Interfacing - More...
4.
Combine a required parameter with the interfaced device. The selection will be
automatically stored in permanent memory.
5. Press
the
Normal Screen
key.
Document No. M1031508
26
Summary of Contents for S/5 Compact Anesthesia Monitor
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 ii ...
Page 10: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 iv ...
Page 60: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 50 ...
Page 61: ...General Service Guide APPENDICES A B Document No M1031508 51 ...
Page 62: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 52 ...
Page 72: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 B 6 6 ...