General Service Guide
1999.
NOTE:
If the DRI level is changed, the monitor will restart automatically.
10. Fill out all necessary documentation regarding the new monitor software.
NOTE: The license agreement of monitor software needs to be in accordance with the monitor
software serial number. Make sure you archive the license agreement in a secure location.
NOTE: The first start-up after software loading takes considerably longer.
3.3.7 Installing the S/5 Network and MemCard Upgrade, U-CMMEM
The following instructions apply to upgrading the monitor with both, the MemCard and Network
options. Detailed upgrade instructions are supplied with the corresponding upgrade product.
NOTE: You can download the option software only onto one monitor. During the downloading of
option software the serial number of the monitor is written on the software card, and if the
downloading for some reason would fail, the software can be downloaded again onto the same
monitor, but not onto any other monitor.
1. Make sure that the monitor is switched to standby.
2. Detach the cover plate on the network connectors at the rear of the frame unit (above the
power supply unit). Leave the collar on the connectors attached.
3. Open the cover for card drive slots (on the left side of the monitor). Insert the option software
card into one of the card drive slots and press the software card firmly in position.
4. Switch the monitor on.
5. Wait until the normal screen appears and check that a message ‘Field upgrade OK’ is
displayed as a note of successful upgrade.
6. Remove the option software card.
7. Enter the service menu and make sure that both options were loaded (
Monitor Setup
–
Install/Service - Service – Frame – Options - ”Net+DC”
)
8. Install the device plate for option software on inside the cover for card drive slots. Close the
cover.
9. Test the monitor with a MemCard and by connecting it onto S/5 Network.
10. Fill out all necessary documentation regarding the upgrade.
3.3.8 Installing the S/5 Network Upgrade, U-CMNET
The following instructions apply to upgrading the monitor with the Network option. Detailed
upgrade instructions are supplied with the corresponding upgrade product.
NOTE: You can download the option software only onto one monitor. During the downloading of
option software the serial number of the monitor is written on the software card, and if the
downloading for some reason would fail, the software can be downloaded again onto the same
monitor, but not onto any other monitor.
1. Make sure that the monitor is switched to standby.
2. Detach the cover plate on the network connectors at the rear of the frame unit (above the
power supply unit). Leave the collar on the connectors attached.
3. Open the cover for card drive slots (on the left side of the monitor). Insert the option software
card into one of the card drive slots and press the software card firmly in position.
4. Switch the monitor on.
Document No. M1031508
17
Summary of Contents for S/5 Compact Anesthesia Monitor
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 ii ...
Page 10: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 iv ...
Page 60: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 50 ...
Page 61: ...General Service Guide APPENDICES A B Document No M1031508 51 ...
Page 62: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 52 ...
Page 72: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 B 6 6 ...