Datex-Ohmeda S/5 Compact Anesthesia and Critical Care Monitor
5. Wait until the normal screen appears and check that a message ‘Field Upgrade OK’ is
displayed as a note of successful upgrade.
6. Remove the option software card.
7. Enter the service menu and make sure that the option was loaded (
Monitor Setup
–
Install/Service - Service – Frame – Options - ”Net”
)
8. Install the device plate for option software on inside the cover for card drive slots. Close the
cover.
9. Test the monitor by connecting it onto S/5 Network.
10. Fill out all necessary documentation regarding the upgrade.
3.3.9 Installing the S/5 Wireless LAN Upgrade, U-CMW
The following instructions apply to upgrading the monitor with the Wireless LAN option. Detailed
upgrade instructions are supplied with the corresponding upgrade product.
NOTE: You can download the option software only onto one monitor. During the downloading of
option software the serial number of the monitor is written on the software card, and if the
downloading for some reason would fail, the software can be downloaded again onto the same
monitor, but not onto any other monitor.
1. Make sure that the monitor is switched to standby.
2. Open the cover for card drive slots (on the left side of the monitor). Insert the option software
card into one of the card drive slots and press the software card firmly in position.
3. Switch the monitor on.
4. Wait until the normal screen appears and check that a message 'Field Upgrade OK' is
displayed as a note of successful upgrade.
5. Remove the option software card.
6. Switch the monitor to standby.
7. Detach the cover for card drive slots and install the antenna cover.
8. Insert the wireless LAN antenna card into the foremost card drive slot.
9. Switch the monitor on.
10. Enter the service menu and make sure that the option was downloaded successfully (
Monitor
Setup -
Install/Service - Service - Frame - Options - "WLAN" or "All"
)
11. Enter the WLAN service menu and enter appropriate Network ID, WEP Algorithm, Key ID and
Encryption Key information (
Monitor Setup
- Install/Service - Service - Frame - Network -
WLAN - WLAN Config
)
12. Check all installed options on the monitor software device plate and attach the Virtual ID
sticker under the device plate at the monitor rear panel.
13. Test the monitor on S/5 Network.
14. Fill out all necessary documentation regarding the upgrade.
Document No. M1031508
18
Summary of Contents for S/5 Compact Anesthesia Monitor
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 ii ...
Page 10: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 iv ...
Page 60: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 50 ...
Page 61: ...General Service Guide APPENDICES A B Document No M1031508 51 ...
Page 62: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 52 ...
Page 72: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 B 6 6 ...