General Service Guide
number 889308, is recommended.
4.7.2 Analog outputs
Direct ECG (X4 pin 7 and X8 pin3)
Delay (max.): 15 ms
Gain ECG (in)/ECG (out): 1 mV/1V
The signals require input impedance of 100 k
Ω
NOTE: The ECG signal is based on the ECG measurement of the Hemodynamic Modules, M-ESTPR
(all variations) and M-NE(12)STPR (all variations). The ECG signal from the modules is channel 1
(ECG1). The channels 2 and 3 (ECG2, ECG3) are not transmitted. Make sure that the signal in
channel 1 (ECG1) is good enough (extensive QRS for IABP synchronization). For further information,
please contact your authorized Datex-Ohmeda distributor.
NOTE: The direct ECG out signal is not available with modules M-ESTP rev. 01, M-EST rev. 00 and
M-ETP rev.
Pressure out (X8 pin 4)
P1 from hemodynamic module
The Invasive pressure output signal is 1 V/100 mmHg, originally ranging from 0 to 300 mmHg, and
with a delay of approximately 25 ms. The signal requires an input impedance of 100 k.
4.8 S/5 Pressure Temp Module, M-PT output signals
The signal output connector on the Pressure Temp Module, M-PT, can be used to interface some
models of IABPs to the S/5 Compact Anesthesia Monitor and S/5 Compact Critical Care Monitor.
The pin assignments are illustrated in table 10. Please contact your local distributor for more
information.
WARNING
Always make sure that the combination complies with the international safety
standard IEC 60601-1-1 for medical electrical systems and with the
requirements of local authorities.
Table 10
Signal output connector pin assignments
Pin Signal
1
2
3
4
ECG out, ECG1
Pressure out, P3
ECG out, ECG1
Ground
1
2
3
4
4.8.1 Analog outputs
The analog output signal specifications are as follows.
ECG (ECG1)
The ECG signal is 1 V/1 mV with a delay of max 15 ms. The signal requires input impedance of 100
k
Ω
.
NOTE: The ECG signal is based on the ECG measurement of the Hemodynamic Modules, M-ESTPR
(all variations) and M-NE(12)STPR (all variations). The ECG signal from the modules is channel 1
(ECG1). The channels 2 and 3 (ECG2, ECG3) are not transmitted. Make sure that the signal in
channel 1 (ECG1) is good enough (extensive QRS for IABP synchronization). For further information,
Document No. M1031508
39
Summary of Contents for S/5 Compact Anesthesia Monitor
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 ii ...
Page 10: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 iv ...
Page 60: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 50 ...
Page 61: ...General Service Guide APPENDICES A B Document No M1031508 51 ...
Page 62: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 52 ...
Page 72: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 B 6 6 ...