Datex-Ohmeda S/5 Compact Anesthesia and Critical Care Monitor
Table 6
Parameters transferred from Dräger Cato, Julian and Narkomed 2C (NAD) to
the S/5 Compact Anesthesia Monitor or S/5 Compact Critical Care Monitor
Selection Waveforms
Numerics
States
Gases / SSS
CO
2
(kPa, mmHg, %)
CO
2
Fi & Et (kPa, %)
O
2
Fi & Et(%)
N
2
O Fi & Et (%)
RR (CO
2
) (1/min)
AA Fi & Et (%)
Tidal Volume insp. (1)
Minute Volume exp. (1)
Ppeak (mbar)
Pplat (mbar); not in Narkomed 2C
PEEP (mbar)
None
SpO
2
Pleth (% full scale)
SpO
2
(%)
Pulse Rate (1/min)
Probe Off
NOTE: The transference of parameters illustrated in tables 5 and 6 represents the maximum. Some
of these parameters (waveforms or numerics) may not be available in all device versions.
NOTE: The transference of waveforms requires the MEDIBUS real-time extension and in Dräger Cato
waveforms are available only on port 1.
4.5 Interfacing printer
It is possible to interface a laser printer (serial or parallel) with the S/5 Compact Anesthesia
Monitor and S/5 Compact Critical Care Monitor via the monitors’ serial port. The printer should be
PCL5 compatible and should contain at least 2MB of memory. Parallel printers require the use of
Serial-to-Parallel Converter, order code 78030.
NOTE: The following information on the printers can be used as a reference when obtaining suitable
printers locally.
WARNING
WARNING
Always make sure that the combination complies with the international safety
standard IEC 60601-1-1 for medical electrical systems and with the
requirements of local authorities.
Connecting the power supply cord of the printer to the wall socket may cause
the printer leakage current to exceed the limit specified for medical equipment.
Always connect the printer to an appropriate separating transformer.
4.5.1 Setting S/5 Compact Anesthesia Monitor or S/5 Compact Critical Care Monitor
interface for printers
1. Press
the
Record/Print
key.
2. Select
Printer Connection
.
3. Select
Serial
.
4. Press
the
Normal Screen
key.
4.5.2 Connection to serial printers
A serial printer is connected to the serial port connector X11 on the frame unit.
Contact your authorized Datex-Ohmeda distributor for advice on suitable serial printers.
Document No. M1031508
34
Summary of Contents for S/5 Compact Anesthesia Monitor
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 ii ...
Page 10: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 iv ...
Page 60: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 50 ...
Page 61: ...General Service Guide APPENDICES A B Document No M1031508 51 ...
Page 62: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 52 ...
Page 72: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 B 6 6 ...