General Service Guide
Table 1
Transference of parameters, Datex-Ohmeda monitors
Device Waveforms
(analog)
Numerics Alarms
Cardiocap
M-INT -> CO
2
Et&Fi Airway gases, Respiration
rate, SpO
2
, Pulse rate
None
Capnomac
Capnomac II
M-INT -> CO
2
, Pleth
Et&Fi Airway gases, Respiration
rate.
None
Capnomac Ultima
M-INT -> CO
2
Et&Fi Airway gases, Respiration
rate, Spirometry,SpO
2
, Pulse rate,
Pleth amplitude
CO
2
, O
2
, Anesthesia agent,
respiration rate, Apnea,
occlusion, SpO
2
, pulse rate.
Normocap 200
Normocap 200 OXY
M-INT -> CO
2
Et&Fi Airway gases, Respiration
rate.
None
Oscar
Oscar II
Oscar OXY
M-INT -> CO
2
, Pleth
Et&Fi Airway gases, Respiration
rate, SpO
2
, Pulse rate, Pleth
amplitude
None
Satlite
Satlite II
Satlite Plus
M-INT -> CO
2
SpO
2
, Pulse rate, Pleth amplitude
None
Satlite Trans
None
SpO
2
, Pulse rate, Pleth amplitude
None
Table 2
Transference of parameters, external monitors
Device Waveforms
Numerics
Alarms
Critikon
Dinamap 1846SX
None NIBP None
Abbott
Oximetrix 3
None SvO
2
/SaO
2
, CO
None
Baxter
Explorer
None C.O.,
SvO
2,
REF, Tblood
None
Baxter
Vigilance
None C.O.,
SvO
2,
C.C.O., Tblood
None
Nellcor
N-100
N-200
N-1000
Pleth (analog)
SpO
2
, Pulse rate
None
4.1.1 Connection to external Datex-Ohmeda monitors
Use the INT-External Device Cable.
1.
Make sure that the power on both monitors is switched off.
2.
Connect the 9 pin D-connector to one of the connectors on the Interface Module, M-INT.
Tighten the finger screws.
3.
Connect the 25 pin D-connector to the corresponding connector on the interfaced monitor.
Tighten the finger screws.
4.1.2 Connection to Critikon Dinamap 1846SX, Abbott Oximetrix 3 and Baxter Explorer
Use the INT-External Device Cable.
1. Make sure that the power on both monitors is switched off.
2. Connect the 9 pin D-connector to one of the connectors on the Interface Module, M-INT.
Tighten the finger screws.
3. Connect the 25 pin D-connector to the corresponding connector on the interfaced monitor.
Tighten the finger screws.
Document No. M1031508
25
Summary of Contents for S/5 Compact Anesthesia Monitor
Page 2: ......
Page 6: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 ii ...
Page 10: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 iv ...
Page 60: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care Monitor Document No M1031508 50 ...
Page 61: ...General Service Guide APPENDICES A B Document No M1031508 51 ...
Page 62: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 52 ...
Page 72: ...Datex Ohmeda S 5 Compact Anesthesia and Critical Care monitor Document No M1031508 B 6 6 ...