General Service Guide
6 GENERAL
TROUBLESHOOTING
Monitor not
functioning
Power cord
connected?
Connect power
cord
Monitor can be
started now?
Batteries were
empty
On/Standby LED
is ON or blinks?
PMB failure.
Replace PMB.
Monitor can be
started now?
Display Unit
failure.
No
Yes
Yes
Yes
No
Possible AC/DC
unit failure.
Replace AC/DC
unit.
No
No
Frame Unit fan is
running?
Yes
No
AC/DC Unit
failure.
Yes
C
M
1_trbl
_
F
low
chart.vsd
Figure 14
Monitor General Troubleshooting Flowchart