MAINTENANCE WITH Q&A
Page 5-4
11. Electrical Panel
Check for components and wires which may vibrate loose during operation.
Check for signs of contactor or relay arcing. Check to see that power and
ground connections remain secure.
12. Operating Current
Check the operating current of the compressors,heaters, circulator and
other components with an Ammeter. Maximum operating values are shown on
the electrical schematic.
13. Check GFCI Receptacle(s) (If Applicable)
Check per instructions on the receptacle. Location of the receptacle(s) is
indicated on the electrical schematic.
14. Tubing Abrasion
Check for evidence of friction wear on all refrigeration lines. Particular care
should be taken in inspecting capillary tubes to pressure gauges/transducers.
Check lines any time the system is shut down.
15. Oil Pressure (Semi-Hermetic Compressors Only)
The oil pressure should run from 15 to 50 PSIG above the suction pressure of
the compressor. A reading below 15 PSIG should be investigated immediately.
16. Humidity Sensor
It is recommended to perform this routine maintenance monthly to avoid
inaccurate readings. If you are having trouble achieving levels or if you can see
dirt, please make sure to clean the sensor.
Depending on the conditions of measurement, the dust
fi
lter should be checked
from time to time. Corroded, discolored, or clogged
fi
lters should be replaced.
17. Product High Limit (Check every 3 months)
Check for proper operation by lowering the setpoint and tripping the safety.
18. Water Strainer (Water Cooled Units)
Check/Clean the water strainer (if equipped).
YEARLY
19. Water-Cooled Condenser (If Applicable)
Annually check the condenser for buildup of scale on the condensing surface.
This interval may need to be shortened if the buildup is of a substantial amount.
20. Chamber Controller Calibration/Veri
fi
cation
Calibrate/Verify the chamber controller settings if all instruments in your facility
are periodically calibrated to one device (metrology), or if a measurement
system component fails. The calibration/veri
fi
cation procedure is located in the
Controller Manual which accompanied your unit.
Voltages dangerous to
life are present in this
equipment.
Electrical
service must be performed
by qualified electricians
only. Failure to follow these
instructions may result in
personal injury and/or loss
of life.
WARNING
!
Summary of Contents for ZP 16
Page 1: ...Installation Operation Maintenance Manual ZP Series 56120 REV AH 10 2019...
Page 4: ...INTRODUCTION THIS PAGE INTENTIONALLY LEFT BLANK...
Page 6: ...CHAMBER LABELS MEANINGS THIS PAGE INTENTIONALLY LEFT BLANK...
Page 40: ...GENERAL DESCRIPTION Page 2 24 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 42: ...INSTALLATION Page 3 2 Figure 1 Transporting a Chamber...
Page 46: ...INSTALLATION Page 3 6 Table 2 Water Usage SI Units...
Page 62: ...OPERATION Page 4 14 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 82: ...MAINTENANCE WITH Q A Page 5 20 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 84: ...SERVICE PARTS WARRANTY Page 6 2 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 88: ...SYSTEM DIAGRAMS Page 9 4 THIS PAGE INTENTIONALLY LEFT BLANK...