OPERATION
Page 4-6
WATLOW F4 CONTROLLER OUTPUTS
Control Outputs
Output IA:
Heating output. Cycles chamber heaters to when heating is called for.
Output IB:
Cooling output. Cycles refrigeration valves when cooling is called for.
The compressors will run as long as cooling is called for before the time-out period has expired.
If cooling is not called for within the preset time-out period, the compressors shut off. They
will turn back when cooling is called for.
Output 2A:
Humidify output. Cycles boiler heater or atomizing nozzle of humidity system when
humidity is called for.
Output 2B:
De-humidify output. Cycles refrigeration valves when dehumidi
fi
ca tion is called for.
The compressors will run as long as de-humidi
fi
cation is called for before the time-out
period is expired. If de-humidi
fi
cation is not called for within the preset time-out period, the
compressors will shut off. The compressors will turn back on when de-humidi
fi
cation is
called for.
Single Set-Point Operation - Humidity (Optional)
1. Turn the humidity system
ON.
2. Use the arrow keys to navigate through the lower display and place the arrow
cursor on setpoint 2.
a. Use the
right arrow key to enter the setpoint
fi
eld.
b. Use the
up and
down arrow keys to change the setpoint.
c. Use the
left or
right arrow keys to exit the setpoint
fi
eld.
3. When returning to temperature only operation, turn the humidity system off.
Failure to do this will result in the unit not reaching low temperatures.
4. Verify that the demineralizer cartridge is fresh. The color of the cartridge should
be dark.
Dry Air Purge
Units Only
D u r i n g h u m i d i t y
operation, the dry air
purge system must be
turned off.
NOTE
Summary of Contents for ZP 16
Page 1: ...Installation Operation Maintenance Manual ZP Series 56120 REV AH 10 2019...
Page 4: ...INTRODUCTION THIS PAGE INTENTIONALLY LEFT BLANK...
Page 6: ...CHAMBER LABELS MEANINGS THIS PAGE INTENTIONALLY LEFT BLANK...
Page 40: ...GENERAL DESCRIPTION Page 2 24 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 42: ...INSTALLATION Page 3 2 Figure 1 Transporting a Chamber...
Page 46: ...INSTALLATION Page 3 6 Table 2 Water Usage SI Units...
Page 62: ...OPERATION Page 4 14 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 82: ...MAINTENANCE WITH Q A Page 5 20 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 84: ...SERVICE PARTS WARRANTY Page 6 2 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 88: ...SYSTEM DIAGRAMS Page 9 4 THIS PAGE INTENTIONALLY LEFT BLANK...