GENERAL DESCRIPTION
Page 2-17
**Performance is based
on 230V, 60 Hz. operation
and a 24°C ambient. For
50 Hz operation, stated
performance and air
fl
ow may
be approximately 17% less.
Speci
fi
cations are subject to
change.
**Electrical requirements
on temperature only units.
Amperage may increase on
humidity models. See quotation
for actual values.
THE PERFORMANCE OF YOUR CHAMBER IS SIGNIFICANTLY
AFFECTED BY THE CHARACTERISTICS OF YOUR TEST
SAMPLE. FACTORS INCLUDING SIZE, MATERIAL, SHAPE,
WEIGHT AND POWER DISSIPATION IF ENERGIZED. THE
TEST SAMPLE SHOULD BE PLACED IN THE CHAMBER IN A
MANNER THAT ALLOWS PROPER VENTILATION. AIR FLOW
IS FROM TOP TO BOTTOM IN THE CHAMBER. TEST SAMPLES
SHOULD BE PLACED ON THE SHELVES, NOT PLACED
DIRECTLY ON THE CHAMBER FLOOR.
PLEASE REFERENCE
YOUR CSZ QUOTATION AS VALUES MAY VARY IN SPECIFIC
CHAMBER PERFORMANCE LISTED IN EACH CHART SEEN
ON PAGES 14 - 17.
NOTE
Table 7. ZP Tundra Speci
fi
cations
Tundra® -45°C to 190°C
Cooling Performance
Cooling
Heating Performance
with Empty chamber in Minutes from:
Rate ²
with Empty Chamber in Minutes from:
24°C
85°C
°C / min
24°C
-34°C
Model
-18°C
-34°C
-40°C
-40°C
93°C
190°C
24°C
ZP-8-2-SCT/AC
6
12
18
30
4.1
10
30
10
ZP-8-3.5-SCT/AC
4
8
12
22
5.6
3.5
12
3.5
ZP-8-6-SCT/WC
3
5
7
12
10.4
3.5
12
3.5
ZP-16-3.5-SCT/AC
5
10
15
25
5.0
10
30
6
ZP-16-6-SCT/WC
3
7
10
18
7.0
10
30
6
ZP-16-10-SCT/WC
2.5
5
7
12
10.4
4
11
3
ZP-32-3.5-SCT/AC
7
15
20
35
3.5
10
35
8
ZP-32-6-SCT/WC
4
9
12
22
5.6
10
35
8
ZP-32-10-SCT/WC
3
6
8
18
7.0
7
20
6
ZP-32-15-SCT/WC
2
4
6
11
11.3
6
16
5
ZP-44-3.5-SCT/AC
11
25
32
52
2.4
20
60
15
ZP-44-6-SCT/WC
5.5
15
20
30
4.1
20
60
15
ZP-44-10-SCT/WC
4
8
12
22
5.6
8
25
7
ZP-44-15-SCT/WC
3
7
10.5
14
9.0
6
15
5
ZP-64-3.5-SCT/AC
14
30
38
65
2.0
25
75
20
ZP-64-6-SCT/WC
7
17
22
38
3.2
25
75
20
ZP-64-10-SCT/WC
5
11
18
24
5.2
10
30
8
ZP-64-15-SCT/WC
4
9
14
18
7.0
8
20
6
ZPS-80-3.5-SCT/AC
16
36
48
75
1.7
30
80
40
ZPS-80-6-SCT/WC
12
24
30
50
2.5
30
80
40
ZPS-80-10-SCT/WC
7
14
25
32
3.9
15
40
12
ZPS-80-15-SCT/WC
5
12.5
19
24
5.2
12
31
10
Summary of Contents for ZP 16
Page 1: ...Installation Operation Maintenance Manual ZP Series 56120 REV AH 10 2019...
Page 4: ...INTRODUCTION THIS PAGE INTENTIONALLY LEFT BLANK...
Page 6: ...CHAMBER LABELS MEANINGS THIS PAGE INTENTIONALLY LEFT BLANK...
Page 40: ...GENERAL DESCRIPTION Page 2 24 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 42: ...INSTALLATION Page 3 2 Figure 1 Transporting a Chamber...
Page 46: ...INSTALLATION Page 3 6 Table 2 Water Usage SI Units...
Page 62: ...OPERATION Page 4 14 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 82: ...MAINTENANCE WITH Q A Page 5 20 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 84: ...SERVICE PARTS WARRANTY Page 6 2 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 88: ...SYSTEM DIAGRAMS Page 9 4 THIS PAGE INTENTIONALLY LEFT BLANK...