MAINTENANCE WITH Q&A
Page 5-1
Table 1. Test Equipment Required
All loads must be removed from the
chamber prior to transport and/or
before maintenance and servicing.
WARNING
!
GENERAL MAINTENANCE
This chapter contains only general maintenance instructions that
can be performed outside the factory. More detailed maintenance
instructions may be found in the supplemental manuals included on
the digital media in the Manual binder that shipped with your unit,
(and, if applicable, the booklets in the rear pocket). You may also
contact the factory for instructions.
WARNING: Only quali
fi
ed maintenance personnel should be
permitted to perform any maintenance procedures.
TEST EQUIPMENT REQUIRED
Table 1 is a list of the test equipment required to test and maintain
the chamber. Equivalent equipment may be used.
Only
qualified maintenance
personnel
should be permitted
to perform any maintenance
procedures.
WARNING
!
NOMENCLATURE
USE
Precision millivolt source
Millivolt source for use in calibrating analog inputs
Type "T" reference table
Reference table used for calibrating temperature inputs
Digital voltmeter
Measures voltage
Ammeter
Measures current
Two sets of manifold refrigeration gauges
with charging hoses
Monitoring refrigeration system pressures and connecting
purge/charge equipment to refrigeration system
One tank of HFC-404A (If Applicable)
Charging refrigeration system
One tank of R-410A (If Applicable)
Charging refrigeration system
One tank of R-508B (If Applicable)
Charging refrigeration system
One tank of R-23 (If Applicable)
Charging refrigeration system
Leak detector
Detecting leaks in refrigeration system
Ohmmeter
Continuity and resistance tests
Digital Clamp-on Ammeter
Reads current draw of electrical components including
heaters, compressors, etc
Summary of Contents for ZP 16
Page 1: ...Installation Operation Maintenance Manual ZP Series 56120 REV AH 10 2019...
Page 4: ...INTRODUCTION THIS PAGE INTENTIONALLY LEFT BLANK...
Page 6: ...CHAMBER LABELS MEANINGS THIS PAGE INTENTIONALLY LEFT BLANK...
Page 40: ...GENERAL DESCRIPTION Page 2 24 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 42: ...INSTALLATION Page 3 2 Figure 1 Transporting a Chamber...
Page 46: ...INSTALLATION Page 3 6 Table 2 Water Usage SI Units...
Page 62: ...OPERATION Page 4 14 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 82: ...MAINTENANCE WITH Q A Page 5 20 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 84: ...SERVICE PARTS WARRANTY Page 6 2 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 88: ...SYSTEM DIAGRAMS Page 9 4 THIS PAGE INTENTIONALLY LEFT BLANK...