crankcase heater is installed on the compressor in order to heat the oil in the crankcase of the
compressor. This helps boil out any refrigerant, so that it can be pumped into the high side of the
system.
If the chamber is equipped with the pumpdown mode of operation, the “Pumpdown” button will be
provided on the Alarm screen if pumpdown is disabled. This provides a manual way to reset
pumpdown mode. Pumpdown will be automatically disabled if main power is off for more than 30
minutes. This is due to the fact that without main power, the crankcase heater will be off and cannot
warm the compressor. Depending on the temperature of the compressor, this could allow refrigerant
to accumulate in the compressor causing a potentially damaging start-up.
The “Pumpdown Disabled” alarm will notify the operator of this condition. The chamber will not
operate when pumpdown is disabled. Pumpdown will automatically reset after the main power has
been on for a period of 4 hours or it can be reset at any time by pressing the “Pumpdown” button.
If your chamber is equipped with pumpdown, main power should remain on at all
times. If power is removed for extended periods of non-use, DO NOT reset pumpdown
after power is applied to the unit. Only silence (reset) the alarm. Allow the 4 hour
warm-up period prior to use or damage to the compressor may result.
2.4.2
Help For Alarms
The EZT-570S provides alarm specific help to assist the operator in determining what caused an
alarm and how it may be corrected. To view the help for an alarm, select the alarm from the alarm list
by touching the alarm’s description. The selected alarm will then be highlighted orange.
Once the alarm is selected, press the help icon in the menu bar. The help screen will then be shown
with information relating specifically to the alarm.