8.9 Configuration
The Configuration utility allows the user to back up the settings of the EZT-570S and save it to a file
for “safe keeping” in case of system damage due to a lightning strike, etc. The utility also allows the
EZT-570S to be configured with a push of a button by selecting from a list of preconfigured setups
that could be created and saved on a USB memory device for import to a new EZT-570S.
Do not import configuration files that are not intended for use with the current
chamber type and hardware configuration of the EZT-570S. Consult our service
department before using the import function. Importing system configurations not
intended for use on your model of chamber may damage to system components and
will void the chamber warranty.
Figure 8-25 Configuration
When the
Export
button is pressed, the EZT-570S configuration data will be written to the USB
memory device. The export file function will use the name entered in the text field to the right of the
export button and create a directory on the memory stick with that name, to contain the configuration
data. The filename can be entered with up to 10 characters which permits multiple configurations to
be identified and saved for later retrieval. The export file will be saved with a name format of
“cfg_
filename
_mm_dd_yy_hh_mm_ss” to identify the configuration according to the date and time it
was created. This prevents multiple configuration files with the same name from overwriting previous
files. It also allows the user to identify a backup of the EZT-570S so that it could be reconfigured to a
previous date if multiple back-ups are created.
The
Import
button is used to write the configuration data to the EZT-570S from the currently loaded
file. To load a configuration file, the USB memory device must first be inserted into the USB port.
When the
Load Configuration
button is pressed, a list of available configuration files from the
memory stick will be shown. Select the desired file from the list and press “Open” to load the file.
Once the file is loaded, the import process can begin. Pressing the “Import” button will begin the
process. When the import is complete, you must cycle power to the EZT-570S in order for the new
configuration data to be loaded into runtime memory for proper operation.
The
Reset default configuration
button can be used to clear the current EZT-570S configuration
and reset the system back to default values. This function clears all user configuration values and
returns the system to an “as new” factory state allowing the system to be configured from scratch.
This function is provided as a means to start over in the event that the system configuration has been
altered to an unknown state and is not operating properly, and a known good configuration is not
available to import.