INSTRUCTIONS FOR USE
This document is the property of
CSO Costruzione Strumenti Oftalmici srl.
Any reproduction, even partial, it is prohibited.
49/56
CAUTION
Danger of material damages. Do not use spray products.
Do not use excessively wet cloths, as they may drip.
If needed, use a damp and well wrung out cloth.
Make sure no liquid penetrates into the device.
Cleaning and disinfection procedures shall be regularly carried out.
Parts of the device that do not come into direct contact with the
patient shall be cleaned at least once a day.
Parts of the device that do come into direct contact with the patient
shall be thoroughly cleaned and disinfected after each application.
The current section described the procedures to be carried out during
use and maintenance, in order ensure proper cleaning and disinfection
of the device and its accessories.
5.2.1
RECOMMENDED PRODUCTS FOR CLEANING AND DISINFECTION
CAUTION
Danger of material damages. Do not use solvents, acidic or basic
solutions (pH <4,5 or >8,0), abrasive or caustic substances, chlorine
-
based and chlorine-derived products.
The Manufacturer is not liable for any damages caused by using
disinfecting products which are not indicated in this manual.
The choice of the most suitable product and procedures for the
cleaning and disinfection of the device takes into account both the
sensitivity of the device to specific substances and the product’s
disinfecting effectiveness.
For cleaning and disinfection procedures, use medical products FDA or
CE approved for medical devices or medical-surgical devices.