| IFU327EN00 - 01/2021
INSTRUCTIONS FOR USE
This document is the property of
CSO Costruzione Strumenti Oftalmici srl.
Any reproduction, even partial, it is prohibited.
39/56
4.4
HOW TO TURN ON THE DEVICE
Read the instructions for use before using the application software.
1
Press the activation switch of the power supply unit on ON.
2
Turn on the PC.
3
Launch the Phoenix application software.
4
Wait until the main screen of the application software is
shown.
5
If this is the first time you’re starting the device, or after a long
period of inactivity, calibration shall be performed. Follow the
instructions given in paragraph
calibration” on page 39.
4.4.1
HOW TO PERFORM DEVICE CALIBRATION
Calibration shall be performed when starting the device for the first
time or after a long period of inactivity of the latter.
The procedure shall be carried out in a dark room so as to simulate
the environmental conditions of a standard capture procedure.
Follow the instructions about calibrating the device which are
given in the Phoenix application software handbook.
Particular attention shall be paid while performing such procedure. It
is important to check device stability before starting with the
procedure.
The calibration is essential to obtain precise measurements.
1
Make sure the calibration tool is clean and undamaged. If
needed, clean it using a soft cloth.
Do not use solvents or diluents to clean the calibration tool.
2
Place the calibration tool on the chin rest.