Immunity test
IEC 60601
-1-2
test level
Conformity
level
Electromagnetic
environment
Electrostatic
discharge. IEC
61000
-4-2
±6 kV contact.
±8 kV air
±6 kV contact.
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
the floors are covered
with synthetic material the
relative humidity should
be at least 30%.
Electrical fast
transient/burst.
IEC 61000
-4-4
±2 kV for
power supply
lines. ±1 kV for
input/output
lines
±2 kV for
power supply
lines. Non-
applicable
Mains power quality shall
be that of a typical
commercial or hospital
environment.
Surge IEC 61000
-
4-5
±1 kV
differential
mode. ±2 kV
common
mode
±1 kV
differential
mode. ±2 kV
common
mode
Mains power quality shall
be that of a typical
commercial or hospital
environment.
Voltage dips.
Short
interruptions and
voltage variations
on power supply
input lines. IEC
61000
-4-11
<5% Un for 0.5
cycles. 40% Un
for 5 cycles.
70% Un for 25
cycles. <5% Un
for 5 s
<5% Un for
0.5 cycles.
40% Un for 5
cycles. 70%
Un for 25
cycles. <5%
Un for 5 s
Mains power quality shall
be that of a typical
commercial or hospital
environment. If the user of
the device requires
continued operation
during power mains
interruptions, it is
recommended that the
device is powered from an
uninterrupted power
supply or battery.
Power frequency
(50/60Hz)
magnetic fields.
IEC 61000
-4-8
3 A/m
3 A/m
Power frequency of the
magnetic fields should be
that of a typical
commercial or hospital
environment.