INSTRUCTIONS FOR USE
| IFU327EN00 - 01/2021
This document is the property of
CSO Costruzione Strumenti Oftalmici srl.
Any reproduction, even partial, it is prohibited.
20
/56
2.8
MANUFACTURER DECLARATIONS
2.8.1
ELECTROMAGNETIC EMISSIONS
The device is designed to be used in a room with the following
electromagnetic characteristics:
Emission test
Compliance Electromagnetic environment
Radio frequency
emission. CISPR 11
Group 1
The device uses radio frequency
energy only for its inner functioning.
The radio frequency emissions of the
device are very low and should not
cause interferences with the near
appliances.
Radio frequency
emission. CISPR 11
Class B
The device can be used in all the
environments, included the domestic
environment. The device can be
connected directly to a low-voltage
power supply net as there is in the
housing units.
Harmonic emissions. IEC
61000
-3-2
Class A
The device can be used in all the
environments, included the domestic
environment. The device can be
connected directly to a low-voltage
power supply net as there is in the
housing units.
Limitation of voltage
changes, voltage
fluctuations and flicker.
IEC 61000
-3-3
Compliant
The device can be used in all the
environments, included the domestic
environment. The device can be
connected directly to a low-voltage
power supply net as there is in the
housing units.