After performing calibration, an examination shall be carried
out by means of the calibration tool (8 mm sphere radius), in
order to check proper calibration of the device.
15
Click on the button NEW PATIENT, enter their personal data,
then confirm and choose the CORNEAL TOPOGRAPHY
examination.
16
After capturing the image, press the EXIT button and process
the captured examination.
17
On the OPTIONS panel, select the curvature’s unit of
measurement in millimetres.
18
Check the value of the reference sphere corresponds to that on
the anterior tangential curvature map.
19
If the elaborated measures are not considered reliable, repeat
the whole calibration procedure.
In case the device is not properly calibrated. Repeat the calibration
procedure.
4.4.2
HOW TO CREATE A NEW PATIENT
1
Click on NEW PATIENT and enter his personal data. If the
patient is already present in the database, you can
automatically search the surname in the surname command
line.
2
A new examination will be created automatically.
3
Select the examination to be performed.
4
The image capture screen will open. Now it will be possible to
capture the image.
4.4.3
HOW TO CREATE A NEW EXAMINATION
1
Click on the button NEW EXAMINATION.
2
Select the examination to be performed.
3
The image capture screen will open. Now it will be possible to
capture the image.