30
Peristaltic Pumps Series IP 100-800
8. Turn the rotor for 180
°
and repeat this
procedure to adjust the second sliding block.
9. Reassemble the housing cover and refill
lubricating fluid (see previous section).
Pressure
Rotor gap X
IP 300
IP 500
0 - 2 bar
28 mm
29 mm
2 - 4 bar
27 mm
29 mm
4 - 6 bar
26 mm
28 mm
6 - 8 bar
26 mm
27 mm
8 - 10 bar
25 mm
26 mm
10 - 13 bar
24 mm
25 mm
Fig. 32
When changing the rotor gap X
make sure that the gap
measurements for both sliding
pieces are identical.
x
Gauge block to adjust
the rotor gap.
5.7.1
Installation of a full fabric hose
Besides the operation with vacuum the pump
can also be operated with a pumping hose
without vacuum bore. The installation
procedure for a so-called full fabric hose is
identical with the procedure for a standard
pumping hose.
Switch off the pump and secure it against
restarting.
Then proceed as described in chapter 5.5.3.
Attention!
Pressure
Rotor gap X
IP 300
IP 500
0-29 psi
1.10 inches
1.14 inches
29-58 psi
1.06 inches
1.14 inches
58-87 psi
1.02 inches
1.10 inches
87-116 psi
1.02 inches
1.06 inches
116-145 psi
0.98 inches
1.02 inches
145-188,4 psi
0.94 inches
0.98 inches