5
Check/re-adjust video levels
On the toolbar, click
Refresh current camera parameters
after you have entered
the target white reference values (see sstep 4).
On the tree view, click
Gain settings
, and then check the
Current camera gain
values
. They should be in a range between 80 and 200:
If a current gain value exceed 200, you could increase the corresponding value at
Analog coarse gain
. If a current gain value is below 80, you could reduce the
corresponding value at
Analog coarse gain
.
If you modified parameters, click
Transfer setting to camera
on the toolbar, and
then start again with step 5.
If you do not achieve satisfying results, you could also adjust values at
Target white
reference values
on the
White control
page (refer to step 4).
On the toolbar, click
Transfer setting to camera
to save the settings non-
permanently on the camera.