Error Messages
5-3
An invalid character string was received. Check to see if you have enclosed the character
string in single or double quotes.
Example : DISP:TEXT “HELLO
-211 Trigger Ignored
A Group Execute Trigger (GET) or *TRG was received but the trigger was ignored. Make sure
that the trigger source should be selected to the bus and the trigger subsystem should be
initiated by INIT[:IMM] command.
-213 Init ignored
An INITiate command has received but could not to execute because of current device
settings.
-221 Settings Conflict
Indicate that the operation can not complete under present states of the power supply. For
example, try to change the output sequence step settings without turning off the output first
will generate this error.
-222 Data out of Range
The value of the numeric parameter for the command is outside the valid range.
Example: VOLT:LEV -3
-224 Illegal Parameter Value
A discrete parameter was received but did not accept by the command. You may have used
an invalid parameter choice. Check the parameter list and choose a correct one.
Example: DISP MAX (MAX is not a valid choice).
-330 Self-test Failed
The complete self-test of the power supply from the remote interface (*TST? Command) failed.
In addition to this error, more specific self-test errors are also reported.
-350 Too many errors
Too many errors occurred (more than 32 errors) and the error queue is overflow. No
additional errors are stored until you remove errors from the queue. Reading the error
message stored in the queue will remove it at the same time. The error queue is cleared when
the power supply has been powered off, or after a *CLS (clear status) command has been
executed.
-410 Query INTERRUPTED
A command which sends data to the output buffer was received, but there were data from a
previous command still in the output buffer (the previous data is not overwritten). The output
buffer is cleared when the power supply has been powered off, or after a *RST (reset)
command has been executed.
-420 Query UNTERMINATED
The power supply was requested to talk (i.e., to send data over the interface) but a command
which sends data to the output buffer has not been received. For example, you may have
executed an APPLy command (which does not generate data) and then attempted an ENTER
statement to read data from the remote interface.
-430 Query DEADLOCKED