Programmable DC Power Supply 62000L Series User’s Manual
4-14
*RCL {0|1|…|15}
This command is used to recall the stored operation states. For more details, please refer to
the “Storing and Recalling Operation States”.
4.2.5 Calibration Commands
The following calibration commands are used to calibrate your power supply. For detailed
command information, please read the command descriptions coming next to the commands.
CALibration
:COUNt?
:CURRent[:DATA] <numeric value>
:CURRent:LEVel {MIN|MAX}
:SECure:STATe {OFF|ON},<code>
:STRing <string>
:STRing?
:VOLTage[:DATA] <numeric value>
:VOLTage:LEVel {MIN|MAX}
CALibration:COUNTt?
This query command is used to return the number of time that it has been calibrated.
CALibration:CURRent[:DATA] <numeric value>
This command is used to enter the CC calibration data. Before using this command, a
calibration level should set first or an calibration sequence interrupt error will be generated.
CALibration:CURRent:LEVel [MINimum|MAXimum]
This command is used to set the CC calibration level of the power supply. There are two
calibration points for the power supply to calibrate CC. The MIN level must be calibrated first
or a calibration sequence interrupt error will be generated.
CALibration:SECure:CODE < code>
This command is used to change the security code. To change the security code, you must
un-secure the calibration first and secure the calibration again with the new code after
entering the new security code with this command. The code is an un-quoted string with 6
alphanumeric characters. No matter the uppercase or lowercase letters.
CALibration:SECure:STATe {OFF|ON},<code>
This command is used to enable/disable the calibration security of the power supply. The
code is an un-quoted string with 6 alphanumeric characters. No matter the uppercase or
lowercase letters.
CALibration:SECure:STATe?
This command is used to query the calibration security state of the power supply. If the
security is enabled, 1 is return. If the security is disabled, 0 is return.
CALibration:STRing <quoted string>
This command is used to write a message about the calibration. For example, you can use
this command to save the calibration date. The string can contains up to 35 alphanumeric
characters and must in an enclosed quote.
CALibration:STRing?
This command is used to query the calibration string.