Programmable DC Power Supply 62000L Series User’s Manual
5-2
There is one or more invalid separator in the command string. You may have used a comma
instead of a colon, semicolon, or blank space; or you may have used a blank space instead of
a comma.
Example: VOLT,10
-108 Parameter not Allowed
More parameters than expected for the command were received. You may have added a
parameter to a command that does not accept a parameter or you entered extra parameters
for the command.
Example: *OPC 1
-109 Missing parameter
Fewer parameters than expected for the command were received. One or more parameters
that are required for the command were omitted.
Example:
VOLT:LEV
-113 Undefined Header
A command that is not valid for this power supply was received. The command may not be a
valid one or you have misspelled it. If you are using the short form of the command, remember
that it the maximum length is four letters.
Example: TRIGG:DEL 3
-121 Invalid Character in Number
An invalid character was found in the number specified for a parameter value.
Example: *ESE B01010102
-124 Too Many Digits
The received numeric parameter was found that contained more than 21 digits.
-131 Invalid Suffix
A suffix was incorrectly specified for a numeric parameter. You may have misspelled the suffix
or the suffix is not available for the parameter.
-138 Suffix not Allowed
The received parameter contains suffix that not accept by the command.
Example : CURR 1V
-141 Invalid Character Data
Invalid character data was found. The character data element may contain an invalid
character or the particular element received was not valid for the header.
-144 Invalid character data length
There are too many characters in the character data.
Example : CAL:SEC:CODE 01234567
-151 Invalid String Data