Programmable DC Power Supply 62000L Series User’s Manual
3-12
damage from high voltage coming from the DUT.
2. Output terminal connection
Connect the Master to the DUT +, the Master output - to the
Slave1 , the Slave1 output - to the Slave2 , and the
Slave2 output - to the DUT - to form the connection in series. The
connection with reversed polarities, such as the Master output
connected to the DUT - and so on, is also available.
3. Master/Slave mode Setup
When the connection in series is used, the output voltage is the sum of
the voltages from the connected power supplies. Each of the individual
power supplies must be properly set in order to obtain the total output
voltage.
The sum won’t be shown on the display. You have to calculate this for
yourself or prepare a load (or a meter) to inspect and confirm the total
output value.
●
Master Unit (Quantity X 1)
:
This setup requires only 1 unit. Before setting this unit to the Master, disable the output.
1. Press the
Menu
button.
2. Rotate the Knob to show MA/SL on the display.
3. Press the
Menu
button to confirm.
4. Rotate the Knob to show SETUP on the display.
5. Press the
Menu
button to confirm.
6. Rotate the Knob to show MODE on the display.
7. Press the
Menu
button to confirm.
8. Rotate the Knob to show S-MA on the display.
9. Press the
Menu
button to finish the settings.
10. To quit the configuration, press any other key to exit.
●
Slave Unit (Quantity X 1 to 6)
:
The buttons used are the same as the ones used for setting the master unit. This connection
requires at least 1 unit, and at most 6 units. For example: If two slave units are required, each
unit has to be set to the slave state. Before setting this unit to Slave, disable the output.
1. Press the
Menu
button.
2. Rotate the Knob to show MA/SL on the display.
3. Press the
Menu
button to confirm.
4. Rotate the Knob to show SETUP on the display.
5. Press the
Menu
button to confirm.
6. Rotate the Knob to show MODE on the display.
7. Press the
Menu
button to confirm.
8. Rotate the Knob to show S-SL on the display.
9. Press the
Menu
button to finish the settings.
10. To quit the configuration, press any other key to exit.
●
STATE ON
:
To activate the Master-Salve function using a series connection, you have to enable the state
from OFF to ON through the following procedures. When the slave state is enabled, the Slave
Slave Unit
Output Connection