Programmable DC Power Supply 62000L Series User’s Manual
3-18
To prevent incorrect actions (Lock) or to return to local operation from the remote control
(Local), the Lock/Local button is used.
The
Lock/Local
button can be used to disable the knob and other buttons so that you can
prevent any unwanted changes. Press this button again to enable the front panel operation. If
active, remote mode usage is also inhibited.
3.10 SEQ (Sequencing Mode)
The 62000L series supports a sequencing mode. This mode allows you to output different
voltages and currents over time. There are 8 steps to the setup. When the SEQ is activated,
the sequence is run from start to finish step-by-step. Before adjusting the sequencing settings,
disable the output.
Note
1. Pressing the V/A button long on the front panel under the SEQ mode in
output condition, the display will show the information of V & A, V & W, or
SEQCNT.
V & A: The display shows the output voltage and current.
V & W: The display shows the output voltage and power.
SEQCNT: The display shows the rest of the SEQ cycle numbers.
2. If you set the start step at step 1 and the stop step at step 4, the power supply
will run in order (i.e. step 1, step 2, step 3, and step 4). If you set the start step
at step 4, and the stop step at step 1, the power supply will still run in order
(i.e. step 4, step 5, step 6, step 7, step 0 and then step 1).
Front Panel Operation:
The next figure of the SEQ family tree of operations on the front panel provides you with an
easy map for setting with the required buttons.
SEQ:
Configuration Sequencing Mode parameter menu
.
You need to disable the output