String parameters contain virtually any set of ASCII characters. A string has to start and end
with matching quotes; either with a single quote or with a double quote. You can type it twice
without any characters in between to include the quote delimiter as part of the string. For
example:
DISP:TEXT <quoted string>
4.2
SCPI Commands
This section collects the commands and provides the descriptions about output setting &
operation, triggering, system, calibration, status, and IEEE-488 conformance.
4.2.1 Output Settings and Operation Commands
This section provides commands to program the power supply output operation. Although the
APPLy
command leaves the most straightforward way to program the CV/CC limit settings,
the
VOLT
and
CURR
commands give more flexibility to change the individual settings.
APPLy {<voltage>|DEF|MIN|MAX}[,{<current>|DEF|MIN|MAX}]
APPLy?
[SOURce:]
CURRent[:LEVel][:IMMediate][:AMPLitude]
{<current>|MIN|MAX}
CURRent[:LEVel][:IMMediate][:AMPLitude]? {MIN|MAX}
CURRent[:LEVel][:IMMediate]:STEP[:INCRement]
{<current>|DEF}
CURRent[:LEVel][:IMMediate]:STEP[:INCRement]? {DEF}
CURRent[:LEVel]:TRIGgered[:AMPLitude] {<current>|MIN|MAX}
CURRent[:LEVel]:TRIGgered[:AMPLitude]? {MIN|MAX}
CURRent:PROTection[:LEVel] {<current>|MIN|MAX}
CURRent:PROTection[:LEVel]? {MIN|MAX}
CURRent:PROTection:STATe {0|1|OFF|ON}
CURRent:PROTection:STATe?
CURRent:PROTection:TRIPped?
CURRent:PROTection:CLEar
VOLTage[:LEVel][:IMMediate][:AMPLitude] {<voltage>|MIN|MAX}
VOLTage[:LEVel][:IMMediate][:AMPLitude]? {MIN|MAX}
VOLTage[:LEVel]:TRIGgered[:AMPLitude] {<voltage>|MIN|MAX}
VOLTage[:LEVel]:TRIGgered[:AMPLitude]? {MIN|MAX}
VOLTage:PROTection[:LEVel] {<voltage>|MIN|MAX}
VOLTage:PROTection[:LEVel]? {MIN|MAX}
VOLTage:PROTection:STATe {0|1|OFF|ON}
VOLTage:PROTection:STATe?
VOLTage:PROTection:TRIPped?
VOLTage:PROTection:CLEar
MEASure
:CURRent[:DC]?
[:VOLTage][:DC]?
:SENSe:EXTernal {OFF|ON}
:SENSe:EXTernal?
OUTPut
[:STATe] {0|1|OFF|ON}
[:STATe]?