8 Maintenance and service
Version 07/2022, Rev. 3.0 of 20/07/2022
• sb
121 / 152
Translation of the original operating manual
Pos : 255 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/080 W artung und I nst andhaltung/ 080-0010- 0091 Kält esyst em Epsilon (nat ürliche, brennbare Kält emitt el) @ 64\m od_1558595727235_68. docx @ 690977 @ 3 @ 1
8.1.8
Refrigeration system
DANGER
Risk of suffocation caused by the refrigerant
When work is performed on the refrigeration system of the freeze-dryer,
refrigerant may escape in the liquid or gas state and under high pressure.
Gaseous refrigerant is heavier than air and high concentration levels of it
may collect on the floor or in pits.
There is a risk of suffocation in the case of high concentration levels.
Possible symptoms are paralysis and unconsciousness. Affected persons
do not notice the fact that they suffocate.
• Only qualified persons are authorised to perform work on the
refrigeration system of the freeze-dryer!
• Ensure good aeration/extraction when working on the refrigeration
system!
WARNING
Risk of cold burns or frostbite caused by the refrigerant
When work is performed on the refrigeration system of the freeze-dryer,
refrigerant may escape in the liquid or gas state and under high pressure.
In the case of skin contact with liquid refrigerant, cold burns or frostbite
may result.
• Only qualified persons are authorised to perform work on the
refrigeration system of the freeze-dryer!
DANGER
Risk of explosion due to refrigerants
The natural refrigerants used are highly flammable and can form an
explosive mixture if their concentration in the ambient air is sufficiently
high.
There is an explosion hazard.
• Work on the refrigeration system of the freeze-dryer must only be
carried out by qualified specialist personnel who have been trained to
handle flammable refrigerants!
• Ensure good ventilation and make sure that no ignition sources (e.g.
soldering iron, welding equipment) are present!
The refrigerant circuit is a closed system. Only certified and qualified
persons are authorised to perform work on the refrigeration system!