9. Zero/Span Adjustment Mode
◆
By measuring scale lower limit (zero side) and higher limit (span side) with your black body furnace and
input black body furnace temperature, this function performs zero/span adjustment.
・
Start zero/span adjustment by push and holding
key,
ENT
key, and
MEASURE
key at the same time
while power OFF.
Status flow is shown on the right.
For each setting and selection, refer to the individual
sections.
Remark
Caution
・
Be sure to prepare a black body furnace and a reference radiation
thermometer.
If those instruments can not be provided, do not perform zero/span
adjustment.
・
Check that emissivity (ratio) is 1.000 then perform the adjustment.
If no key operation has done for 30sec., power
supply is turned OFF automatically.
(IR-HAI/IR-HAQ)
Flow Chart
Initialize Zero/Span
InGaAs Zero Adjustment
Perform Zero/Span
Adjustment
InGaAs Span Adjustment
Si Zero Adjustment
Si Span Adjustment
2-Color Zero Adjustment
2-Color Span Adjustment
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
(IR-HAI/IR-HAQ)
(IR-HAS/IR-HAQ)
(IR-HAS/IR-HAQ)
(IR-HAQ)
(IR-HAQ)