INST.No.INE-567-P1CE
-42-
15. List of Parameters
Name
Parameter Name
Sub
display
Setting range
Initial value
Section
of this
manual
E
m
is
siv
ity
(r
at
io
)
Emissivity (ratio) setting
Single-color: 1.900 to 0.100
2-color: 1.200 to 0.800
1.000
5.1
S
et
tin
g
m
od
e
Lower limit temperature alarm
AL
"
oFF" or
300 to 1000°C (IR-HAI)
600 to 2000°C (IR-HAS)
400 to 3000°C
(IR-HAQH)
oFF
6.1
Higher limit temperature alarm
AH
6.2
Signal modulation*1
Modu
dELy/PEAk
dELy
6.3
Modulation time constant
tAu
0.0, 0.2, 0.5, 1.0sec.
0.0sec.
6.4.1
Attenuation Rate
dEC
0, 2, 5, 10°C/sec.
0°C/sec.
6.4.2
Measuring unit selection
unit
C, F
C
6.5
Thermocouple measurement
tC
oFF/on
oFF
6.6
2-color type/Single-color wide
range type*2
CoLr
2, 1
2
6.7
Memory mode
mmod
mAn/int
mAn
6.8
Memory interval
int
1 to 7200sec.
60sec.
6.9
Data saving folder
FLdr
0000 to 9999
0000
6.10
Data saving file
FiLE
0000 to 9999
0000
6.11
Date/Time
yEAr
2016 to 2099
2016
6.12
mo
1 to 12
1
dAy
1 to 31
1
Hour
0 to 23
0
min
0 to 59
0
SD card data initialization
Frmt
no/yES
no
6.13
Z
er
o
・
S
pa
n
ad
ju
st
m
en
t
m
od
e
Initialize zero/span
adjustment
init
oFF/on
oFF
9.1
Perform zero/span adjustment
CAL
no/yES
no
9.2
InGaAS zero adjustment
InGZ
0 to 9999
300 (IR-HAI)
400 (IR-HAQH)
9.3
InGaAS span adjustment
InGS
0 to 9999
1000 (IR-HAI)
600 (IR-HAQH)
9.4
Si zero adjustment
Si Z
0 to 9999
600 (IR-HAS,
IR-HAQH)
9.5
Si span adjustment
Si S
0 to 9999
2000 (IR-HAS)
3000 (IR-HAQH)
9.6
2-color zero adjustment
2C Z
0 to 9999
600 (IR-HAQH)
9.7
2-color span adjustment
2C S
0 to 9999
2000 (IR-HAQH)
9.8
*1: The parameters of modulation ratio selection differ by the signal modulation mode.
Refer to "6.3 Single Modulation Selection" for details.
*2: 2-color/single-color wide selection is displayed in Model IR-HAQ only