INST.No.INE-567-P1CE
-23-
7.1 Standard Measuring Mode
1) Sight through the viewfinder and match the center circle
of the targeting circle to the center of object measured.
2) Turn ON the power by holding down
MEASURE
key.
3) Start measurement by pushing
MEASURE
key. Status
marker "MEAS" lights and temperature is displayed on the
main display.
Measurement continues while holding down on the key.
4) When
MEASURE
key is released, status marker "MEAS"
turned OFF and status marker "HOLD" lights,
measurement is paused and holds measured value.
5) By pushing
SEL
key, the sub marker switches "MAX"
(highest temperature)
→
"AVE" (average temperature)
→
"TC" (thermocouple temperature)
→
"MIN" (lowest
temperature) and each values can be displayed.
Remark
If the measured value is +21°C or more of measuring
range higher limit, "oFL" is displayed, and -20°C or
less of measuring range lower limit, "uFL" is
displayed.