20-5
6. Accuracy Ratings
No.
Input
type
Measuring range
Reference
range
Accuracy
rating
01
-13.80 to 13.80mV
±
13.8mV
02
-27.60 to 27.60mV
±
27.6mV
03
-69.00 to 69.00mV
±
69.0mV
04
-200.0 to 200.0mV
±
200.0mV
05
DC
(mV)
-500.0 to 500.0mV
±
500.0mV
06
-2.00 to 2.00V
±
2V
07
-5.00 to 5.00V
±
5V
08
-10.00 to 10.00V
±
10V
09
-20.00 to 20.00V
±
20V
10
DC
(V)
-50.00 to 50.00V
±
50V
±
0.1%
±
1d
21
-200.0 to 300.0°C
±
13.8mV
22
-200.0 to 600.0°C
±
27.6mV
23
K
-200 to 1370 °C
±
69.0mV
24
-200.0 to 200.0°C
±
13.8mV
25
-200.0 to 350.0°C
±
27.6mV
26
E
-200 to 900°C
±
69.0mV
27
-200.0 to 250.0°C
±
13.8mV
28
-200.0 to 500.0°C
±
27.6mV
29
J
-200 to 1200 °C
±
69.0mV
30
-200.0 to 250.0°C
±
13.8mV
31
T
-200.0 to 400.0°C
±
27.6mV
32
0 to 1200 °C
±
13.8mV
33
R
0 to 1760 °C
±
27.6mV
34
0 to 1300 °C
±
13.8mV
35
S
0 to 1760 °C
±
27.6mV
36
B
0 to 1820 °C
±
13.8mV
±
0.1%
±
1d
37
-200.0 to 400.0°C
±
13.8mV
38
-200.0 to 750.0°C
±
27.6mV
39
N
-200 to 1300 °C
±
69.0mV
40
W
-WRe26 0 to 2315 °C
±
69.0mV
±
0.15%
±
1d
No. Input type
Measuring range
Reference
range
Accuracy
rating
41
WRe5
-WRe26 0 to 2315 °C
±
69.0mV
43
PtRh40
-PtRh20 0 to 1888 °C
±
13.8mV
44
-50.0 to 290.0°C
±
13.8mV
45
-50.0 to 600.0°C
±
27.6mV
46
NiMo-Ni
-50 to 1310 °C
±
69.0mV
47 CR-AuFe 0.0 to 280.0K
±
13.8mV
±
0.2%
±
1d
48
0.0 to 350.0°C
±
13.8mV
49
0.0 to 650.0°C
±
27.6mV
50
Platinel
Ⅱ
0 to 1395 °C
±
69.0mV
51
-200.0 to 250.0°C
±
13.8mV
52
-200.0 to 500.0°C
±
27.6mV
53
U
-200.0 to 600.0°C
±
69.0mV
±
0.15%
±
1d
54
-200.0 to 250.0°C
±
13.8mV
55
-200.0 to 500.0°C
±
27.6mV
56
L
-200 to 900.0°C
±
69.0mV
±
0.1%
±
1d
70
-140.0 to 150.0°C
160
Ω
±
0.15%
±
1d
71
-200.0 to 300.0°C
220
Ω
72
Pt100
(JIS’97)
-200.0 to 850.0°C
400
Ω
±
0.1%
±
1d
73
-140.0 to 150.0°C
160
Ω
±
0.15%
±
1d
74
-200.0 to 300.0°C
220
Ω
75
Pt100
QPt100
JIS’89
-200.0 to 649.0°C
400
Ω
±
0.1%
±
1d
76
-140.0 to 150.0°C
160
Ω
±
0.15%
±
1d
77
-200.0 to 300.0°C
220
Ω
78
JPt100
-200.0 to 649.0°C
400
Ω
±
0.1%
±
1d
79
Pt50
-200.0 to 649.0°C
220
Ω
±
0.1%
±
1d
80
Pt-Co
4.0 to 374.0
K
220
Ω
±
0.15%
±
1d
Note) "
±
1 d" at accuracy rating stands for
±
1 digit
Note)Thermocouple inputs do not include reference junction compensation accuracy.
The Indication equivalent to 200
μ
V may vary under the test environment of EMC directive.
Exceptions to accuracy rating
Input type
Measuring range
Accuracy rating
K,E,J,T,L
-200 to 0°C
R,S
0 to 400°C
±
0.2%
±
1digit
0 to 400°C
Not specified
B
400 to 800°C
±
0.15%
±
1digit
N,U
-200 to 0°C
±
0.3%
±
1digit
0 to 100°C
±
4%
±
1digit
W-WRe26
100 to 300°C
±
0.5%
±
1digit
0 to 300°C
±
1.5%
±
1digit
PtRh40-PtRh20
300 to 800°C
±
0.8%
±
1digit
0 to 20
K
±
0.5%
±
1digit
CR-AuFe
20 to 50
K
±
0.3%
±
1digit
Pt100(’97)
700 to 850°C
±
0.15%
±
1digit
Pt-Co
4 to 50
K
±
0.3%
±
1digit
Note) Thermocouple input is converted into the reference range.