28
Vectra® Neo Clinical Therapy System
SYSTEM SPECIFICATIONS
POWER (COMBINATION AND
ELECTROTHERAPY UNITS)
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 - 240 V AC, 2.5A to 1.25A, 50/60 Hz
Electrical Class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CLASS I
Mode of Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous
Electrical Type (Degree of Protection)
Ultrasound . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TYPE B
Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TYPE B
Electrotherapy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TYPE BF
Electrotherapy & sEMG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TYPE BF
Ultrasound & Electrotherapy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TYPE B
NOTE:
All waveforms except High Voltage Pulsed Current (HVPC) have been
designed with a 200 mA current limit. VMS™, VMS™ Burst ,and all TENS
waveform output intensities are measured, specifi ed, and listed to peak, not
peak to peak.
SYSTEM SPECIFICATIONS AND DIMENSIONS
Width
Depth
Height
Weight
Module
11.12” (28.2448 cm)
6.34” (16.1036 cm)
1.43” (3.6322 cm)
1lb (0.453592 kg)
Head @ 45 degree with Base (Tabletop)
15.89” (40.3606 cm)
15.89” (40.3606 cm)
22.05” (56.007 cm)
20.7lb (9.389362 kg)
Cart Lowered (with casters)
23.94” (60.8076 cm)
26.19” (66.5226 cm)
27.41” (69.6214 cm)
29.4lb (13.33562 kg)
Cart Raised (with casters)
23.94” (60.8076 cm)
26.19” (66.5226 cm)
30.15” (76.581 cm)
Head and raised cart with screen @ 90deg
23.94” (60.8076 cm)
26.19” (66.5226 cm)
52.85” (134.239 cm)
48.9
lb (22.18067 kg)
GENERAL SYSTEM OPERATING AND STORAGE
TEMPERATURE
Operating Conditions
The device will meet its requirement under the following conditions:
Temperature: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10° C to 45° C
Relative Humidity: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0% to 90%
Atmospheric Pressure: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 700hPa to 1060hPa
Transport and Storage Conditions
The device will remain in proper condition under the following conditions:
Temperature: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Above 0° C freezing to +60°C
Relative Humidity: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . max 95%
Atmospheric Pressure: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 700hPa to 1060hPa
Summary of Contents for Vectra Neo
Page 1: ...Vectra Neo Clinical Therapy System User Manual Operator and Installation Instructions Rx Only ...
Page 2: ......
Page 5: ......
Page 85: ......