77
Vectra® Neo Clinical Therapy System
APPENDIX 2
Recommended separation distances between
portable and mobile RF communications equipment and the Vectra® Neo Clinical Therapy System
The Vectra® Neo Clinical Therapy System is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Vectra® Neo Clinical Therapy System can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Vectra® Neo Clinical Therapy System as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output
power of transmitter
P
(W)
Separation distance according to frequency of transmitter
d
(m)
150 kHz to 80 MHz
d
=
[
3,5
]]
√
P
V
1
(where
V
1
= 3V)
80 MHz to 800 MHz
d
=
[
3,5
]]
√
P
E
1
(where
E
1
= 3V/m)
800 MHz to 2,5 GHz
d
=
[
7
]]
√
P
E
1
(where
E
1
= 3V/m)
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is aff ected by absorption and refl ection from structures,
objects and people.