80
Vectra® Neo Clinical Therapy System
MAXIMUM CURRENT OUTPUT CALCULATIONS BASED ON 250 MW/CM
2
POWER DENSITY LIMIT
APPENDIX 3
Electrode Part
Number
Dimensions (inch)
(shape)
Area (cm
2
)
Maximum Power
Density (mW/cm
2
)
Maximum RMS
Current Based on
Power Density (mA)
Maximum RMS
Current Allowed by
System (mA)
42045
5 × 2.75 (rectangle)
88.71
250
210.61
90.5
42159
2.5 × 1.5 (ellipse)
19.00
250
97.47
90.5
42160
4 × 2 (ellipse)
40.54
250
142.37
90.5
42170
1.25 (circle)
7.92
250
62.92
62.92
42171
2 (circle)
20.27
250
100.67
90.5
42172
2.5 × 1.5 (ellipse)
19.00
250
97.47
90.5
42173
5 × 3 (ellipse)
76.01
250
194.95
90.5
42174
2 × 2 (square)
25.81
250
113.60
90.5
42175
3.5 × 2 (rectangle)
45.16
250
150.27
90.5
42176
2 × 2 (square)
25.81
250
113.60
90.5
42177
3.5 × 2 (rectangle)
33.87
250
130.13
90.5
42178
2 × 2 (square)
25.81
250
113.60
90.5
42179
3.5 × 2 (rectangle)
45.16
250
150.27
90.5
42181
3.5 × 2 (rectangle)
45.16
250
150.27
90.5
42182
2 (circle)
20.27
250
100.67
90.5
42183
2 × 2 (square)
25.81
250
113.60
90.5
Summary of Contents for Vectra Neo