45
Vectra® Neo Clinical Therapy System
5. US
Head
Warming
Select the US Head Warming icon to turn on or off
the ultrasound warming feature during ultrasound
treatment setup. The available choices are either On
or Off. The default setting is Off.
6. E-Stim Contact Quality
Select the E-Stim Contact Quality icon to turn on or
off the electrotherapy contact quality monitoring
function. The available choices are either On or Off.
The default setting is On, and in the On position, the
unit will monitor the contact quality of the electrodes
on all waveforms, excluding High Voltage Pulsed
Current (HVPC) and Microcurrent.
NOTE:
If the impedance feedback exceeds 5000 ohms
(tolerance
±
20%) and the set intensity is ≥ 10 mA CC
or 10 V CV, then the unit will display an information
message and set the intensity to zero for all channels
associated with this treatment.
7. Restore Default Unit Settings
Select the Restore Default Unit Settings icon to reset
all of the following settings back to their factory
defaults:
• Volume
• LCD
Brightness
• US
Coupling
• US
Head
Warming
8. Restore Default Protocols
Select the Restore Default Protocols icon to reset all
protocols (factory, custom, and the 5 custom protocol
shortcuts) to their factory defaults.
9. Display Unit Version Information
Select the Display Unit Version Information icon to
view the install status of each module. For all installed
modules, the PCB information and the Software
version for that module are displayed.
If a valid software upgrade USB flash drive is inserted
into the unit, the version of software for each module
that is on the USB flash drive is displayed.
NOTE:
USB flash drive version is displayed in yellow
if it is newer than the current software version for the
applicable module.
10. Export Protocols to USB flash drive
Select the Export Protocols to USB flash drive icon to
export all protocols (both default and custom) to a
valid USB flash drive.
UTILITIES AND OPTIONS (CONTINUED)
11. Import Protocols from USB Flash Drive
Select the Import Protocols from USB Flash Drive icon
to import all protocols from a valid USB flash drive.
Any protocols with the same name as an existing unit
protocol will be overwritten by the imported protocol,
and any protocols of a different name on the unit will
remain intact.
12. Date and Time
Select the Date and Time icon to set the date and time
on the unit.
13. Clinic Name
Select the Clinic Name icon to enter the name of your
clinic. The clinic displays on the Home screen and on
the patient treatment reports saved to the USB flash
drive. The following icons will perform the functions
listed:
• Left:
Moves the cursor to the left one space per
instance pressed.
• Right:
Moves the cursor to the right one space per
instance pressed.
• Space:
Inserts one space per instance pressed.
• Clear:
Clears the entire entry.
• Delete:
Once the cursor has been moved using the
Left or Right icons, press the Delete icon to delete
the character.
• Caps:
Press the Caps icon to capitalize the letters or
to make letters lowercase.
• Cancel:
Press the Cancel icon to cancel the clinic
name entry function and return to the Utilities
screen.
• Accept:
Press the Accept icon once entry of the
clinic name is complete. The clinic name should
now appear in the title bar area of the screen.