51
Vectra® Neo Clinical Therapy System
DEVICE USER INTERFACE
The Vectra® Neo Clinical Therapy System Ultrasound
modality allows the user to select specifi c Applicator
recommendations and edit treatment parameters for
various syndromes requiring the use of ultrasound
therapy. The following information gives general
instructions for the setup of Ultrasound Therapy when
selecting Ultrasound from the Home screen.
WARNING
• The
Vectra
®
Neo Clinical Therapy System may be susceptible to
Electrostatic Discharge (ESD) at greater than ±4 kV when first grasping
either the Ultrasound or Laser Applicator. In the event of such a
discharge, the Vectra
®
Neo Clinical Therapy System may experience
communication loss with the installed modules. The Vectra
®
Neo
Clinical Therapy System will terminate all active outputs (stim,
ultrasound, laser), automatically place the unit in a safe state, and
issue an error message 301 or 307.
-
To recover from an error message 301 or 307, turn the unit off and
on using the ON/OFF switch located at the top of the display. Once
the system restarts, re-initiate all treatments that were
interrupted.
• To prevent Electrostatic Discharge (ESD) at greater than ±4 kV:
-
Grasp and hold the Ultrasound or Laser Applicator prior to starting
treatment. If the applicator must be put down prior to completion
of treatment, stop the current treatment first and then place the
applicator in the holder.
-
Maintain humidity in the use environment to at least 50% relative
humidity.
-
Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, DJO recommends implementing
additional controls to maintain relative humidity to at least 50%.
-
Communicate these ESD-precautionary procedures to healthcare
staff, contractors, visitors, and patients.
NOTE:
Prior to starting treatment, replace the ultrasound
faceplate on the side of the module not being used with a
blank faceplate.
ULTRASOUND OPERATION
Complete the following steps to begin Ultrasound
treatment:
1. Touch the screen to activate the system. Read and
carefully follow the instructions on the screen.
• To prepare the patient’s skin for Ultrasound Therapy,
prepare patient as described in the ULTRASOUND
PATIENT PREPARATION section on page 41.
2. From the Home screen, select the Ultrasound icon. The
screen shown below will appear:
• To view information explaining the modality, select
the Modality Description icon. Press the Up and
Down icons to view additional text. Press the Back
Arrow icon to return to the previous screen or the
Home icon to return to the Home screen.
• To view applicator information, press the Applicator
Information icon:
- Touch the body part you wish to treat.
- Information about applicator sizes, frequency,
and treatment directions will appear as seen on
the screen above. Press the Down icon to scroll
through the information.
- Press the Back Arrow icon to return to the
previous screen or the Home icon to return to the
Home screen.
• Use the Intensity dial to set therapy intensity:
- Clockwise – Increases Intensity
- Counterclockwise – Decreases Intensity