Intelect® Shortwave 400
50
mAiNTENANCE
Electrical Test (Acc. to IEC 601-1)
(12 month intervals or as dictated by you Health Care Facility)
• Earth leakage current < 0,5 mA (normal condition)
• Earth leakage current < 1,0 mA (single fault condition)
• Housing leakage current < 0,1 mA (normal condition)
• Housing leakage current < 0,5 mA (single fault condition)
• Protective conductor resistance < 100 mΩ
NOTE:
It is the responsibility of the Health Care Facility to verify that the unit complies with the facility, local and national Earth
Leakage limits.
NOTE:
We recommend that the safety inspections be entered in a medical products log in order to document the results of the
inspection.
If the unit is not safe for operation, then it must be
repaired by the authorized service personnel and the
operators must be informed of the dangers posed by
the unit.