Intelect® Shortwave 400
44
OPErATiON
BASiC OPErATiON (CONTiNuEd)
Beginning the Treatment
Treatment is always started from the basic menu by turning up the output power by means of the multi-function rotary switch
from zero to the desired value. The treatment can be interrupted at any time by pressing the STOP key.
type of Electrode
Max . Effective Power
Diode
120 Watt
Diplode
200 Watt
Capacitive Plate Electrode Ø 80 mm
80 Watt
Capacitive Plate Electrode Ø 120 mm
200 Watt
Capacitive Plate Electrode Ø165 mm
400 Watt
Soft Rubber Electrode 120 x 180 mm
250 Watt
Soft Rubber Electrode 145 x 250 mm
400 Watt
Before beginning treatment, you must read and observe the safety instructions and the list of contraindications.
Refer to
Safety Instructions, Personal Safety and Contraindications
for operational safety information.
In order to prevent excessive warming of tissue, the
effective output power, as summarized in the
Type of
Electrode/Max. Effective Power
table must not be
exceeded.
DANGER