Intelect® Shortwave 400
49
mAiNTENANCE
Safety Inspections
The following safety inspections must be performed on this unit. This must be done by persons who, based on training,
knowledge or practical experience, are capable of conducting the inspections correctly and independently.
Visual Inspection (Daily)
• Housing not deformed?
• Power Cable undamaged?
• Electrode connection sockets undamaged?
• Power switch OK?
• Electrode Cable undamaged?
• Capacitive Plate Electrodes, Diplodes, Diodes undamaged (no fissures or brittle material)?
• Soft Rubber Electrodes undamaged (no thin areas or holes)?
Functional Test
(Daily)
• Correct function of indicators
• Display of operating modes
• Time setting, check timer for accuracy, (e.g. ± 5 s)
• Verify power output using a neon indicator
mAiNTENANCE ANd rEPAirS (CONTiNuEd)