Intelect® Shortwave 400
46
OPErATiON
Changing the Basic Settings
Switch off the power switch.
Press the multi-function rotary switch and hold it down.
Switch the unit on by means of the power switch and wait (with the multi-function rotary switch depressed) until the
basic settings menu is displayed.
Turn the multi-function rotary switch until the desired parameter line is selected.
Press the multi-function rotary switch. The value to be changed in this line is displayed inversely.
Set the desired parameter value by turning the multi-function rotary switch.
Confirm the set value by pressing the multi-function rotary switch. The entire line is then inverted again.
Change further basic parameters or switch to the main menu by pressing the function key (ESC).
Parameter
Default value
Range
Pulse duration
400 µs
200 - 600 µs
Volume
50 %
0 – 100 %
Service code*
-
-
LCD language**
any
E, F, S
LCD brightness
70 %
10 - 100 %
LCD contrast
70 %
0 - 100 %