Intelect® Shortwave 400
SPECiFiCATiONS
30
Main Menu
Main menu after unit is switched on
5. Function key for activation of the output power
parameter
6. Function key for activation of the treatment duration
parameter
7. Function key for activation of the operating mode
parameter
8. Function key for setting the frequency in pulsed mode
9. Function key for selection of the indications menu
10. Function key for selection of the programming menu
11. Display fields for identifying the key functions
12. Text output display field
13. Matching display field
14. Effective power display field (patient effective power)
Parameters can be set in three ways:
1. Direct input of parameter with the function keys
(5 through 8)
2. Selection of a disorder in the indications menu (9)
3. Selection of a program saved in the programming
menu (10)
The treatment is always started by turning up the output
power.
Figure 9
Intelect Shortwave 400 Main Menu
1. Output power display field (amplitude)
2. Duration of treatment display field
3. Operating mode display field (pulsed or continuous
mode)
4. Pulse frequency in pulsed mode display field
PrOGrAm dESCriPTiON (CONTiNuEd)