Intelect® Shortwave 400
SAFETy iNSTruCTiONS
11
• Clean and disinfect the unit only when the Mains
Power Supply is deactivated (power switch off, power
plug disconnected).
• Clean and disinfect the unit only by means of
disinfection by wiping. Disinfecting by spraying can
damage the unit due to penetrating moisture.
• Never clean the unit with abrasives, disinfectants or
solvents that could scratch the housing or damage the
unit.
• Never perform unauthorized service work.
• All service work must be performed only by service
technicians who have been authorized by the
manufacturer.
Do not connect the unit to the Mains Power Supply until
these requirements have been fulfilled:
• Before putting the unit into operation, check to make
sure that the electrode connection cable and the
electrodes are undamaged and have been connected
correctly to the unit.
• Never operate the unit with open outputs, (i.e. without
electrodes).
• Do not operate the unit for an extended period with
no load (without a patient), especially in coil (induction
field) mode. When operating the unit without power
output, induction field electrodes could be destroyed
due to overheating.
• Pay attention to the routing of the electrode
connection cables. These must always be in the air and
must never lie on surfaces.
• Keep chip cards, magnetic cards, audio and video
cassettes and other data media that are susceptible to
interference away from the unit.
PrOTECTiON OF ThE uNiT (CONTiNuEd)