74
V-ACTOR HF HANDPIECE OPERATIONS, SET-UP, CLEANING, AND MAINTENANCE
EN
INTELECT
®
RPW 2 USER MANUAL
TROUBLE-SHOOTING
ACCESSORIES AND SPARE PARTS
TECHNICAL SPECIFICATIONS
Subject to technical changes
Note: In the event of the medical product being transferred to third parties, the following must be observed:
– The complete device documentation must be delivered together with the medical product.
– The medical product may only be exported to a foreign country if the medical product and the corresponding indications are allowed there.
This device complies with the applicable standards.
For information about conformity with directives, please refer to the separate operating manual for your control device.
*V-Actor is a registered trade mark of Storz medical AG
Fault description
Possible cause
Corrective actions
No compressed air supply
Leaks in handpiece cable or cable not properly connected
Check the cable and tube connections and replace them, if
necessary
No power output
Handpiece defective
Replace the handpiece
Accessories
Art. no.:
V-ACTOR HF
®
handpiece set
19365
25 mm vibration transmitter
28740
40 mm vibration transmitter
28741
10 mm spherical vibration transmitter without vibration transmitter screw cap
29742
V-ACTOR handpiece
V-ACTOR operating frequency
1 – 50 HZ
Energy selection
in steps from 1 – 5 bar
Ambient temperature during operation
10 °C – 30 °C
Ambient temperature during storage and transport
0 °C – 60 °C
Ambient air pressure during operation
800 – 1060 HPA
Ambient air pressure during storage and transport
500 – 1060 HPA
Air humidity during operation
5 – 55%, non-condensing
Air humidity during storage and transport
5 – 95%, non-condensing
Weight incl. cable, filled
approx. 400 G
Protection against ingress of water
IPX0
Summary of Contents for Intelect RPW 2
Page 32: ...32 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Select transmitter ...
Page 50: ...50 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Delete one session ...
Page 55: ...55 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Delete one protocol ...
Page 61: ...61 USER INTERFACE EN INTELECT RPW 2 USER MANUAL 4 Full screen image Close full screen mode ...