68
V-ACTOR HF HANDPIECE OPERATIONS, SET-UP, CLEANING, AND MAINTENANCE
EN
INTELECT
®
RPW 2 USER MANUAL
OPERATION
The handpiece is operated via the display on the Intelect
®
RPW 2. The settings can either be made manually on the display or
preprogrammed settings can be applied.
START-UP
• Connect the V-ACTOR HF
®
handpiece to the Intelect RPW 2.
• Set the pulse energy in V-ACTOR HF operating mode to an initial value of 2 bar.
• Activate the trigger button.
Note: The trigger button functions as an on/off switch when it is pressed briefly. Pressing it for longer causes it to function as a tip switch, i.e. the
pulses will continue until the button is released.
FUNCTIONAL CHECKS
Perform the following functional checks after the device has been installed:
• Set the energy level in V-ACTOR HF mode to 2.4 bar.
• Reset the actual number of pulses on the display of the control panel.
• Release pulses with a pulse frequency of 30 Hz.
• Check that the triggered pulses are correctly counted on the treatment pulse counter of the control device.
STANDARD SETTINGS
• Before each treatment, make sure that the pulse counter is set to zero.
• Start the V-ACTOR HF treatment at an energy level of 2 bar and a frequency of 20 Hz.
• There is a relationship between pressure and frequency; when one increases the other decreases. The V-Actor HF can not be
used at full frequency and full pressure at the same time. Please find the chart showing the relationship between pressure and
frequency on page 70.
*V-Actor is a registered trade mark of Storz medical AG
Attention!
Malfunction of the device or its components is possible!
Before starting treatment, it is essential to perform the FUNCTIONAL CHECKS described below.
Summary of Contents for Intelect RPW 2
Page 32: ...32 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Select transmitter ...
Page 50: ...50 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Delete one session ...
Page 55: ...55 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Delete one protocol ...
Page 61: ...61 USER INTERFACE EN INTELECT RPW 2 USER MANUAL 4 Full screen image Close full screen mode ...