CLEANING, MAINTENANCE, OVERHAUL
EN
INTELECT
®
RPW 2 USER MANUAL
1. Disconnect the handpiece from the control device.
• Unscrew the shock transmitter screw cap
respectively the clutch for the spine and fasciae
shock transmitters from the handpiece.
2. Unscrew the shaft from the handpiece and pull it out of
the handpiece handle.
• Use the supplied open-end spanner (3) for this
purpose.
3. Pull the tightly fitting guide tube out of the shaft. If
necessary, use a thin metal rod or the supplied hexagonal
spanner as a pulling tool by inserting it through the
openings in the guide tube.
4. A corresponding fixture is provided in the handpiece
handle to hold back the projectile. To remove the projectile,
hold the handpiece handle with its opening pointing down.
Gently knock the handle against the table surface until the
projectile falls out. In the event that the projectile breaks
apart due to overloading, a fragment of the projectile may
be left behind inside the guide tube.