77
FALCON HANDPIECE CLEANING, MAINTENANCE; OVERHAUL
EN
INTELECT
®
RPW 2 USER MANUAL
FALCON
®
HANDPIECE CLEANING, MAINTENANCE; OVERHAUL
CLEANING
Regular cleaning ensures perfect hygiene and operation of the Falcon
®
handpiece. The handpiece, in particular the shock
transmitter, must be thoroughly cleaned and disinfected after each therapy session.
• Clean the coupling gel off the handpiece using a cleaning agent that is suitable for surface cleaning.
• Disinfect the handpiece with an alcohol-based disinfectant that is suitable for surface cleaning.
Component
Procedure
Interval
Handpiece shaft and cushion
clean and disinfect
daily
or after 20,000 shocks (whichever comes first)
Guide tube
clean from inside with brush
every day
Shock transmitters and O-rings
clean in ultrasonic bath and disinfect
after each treatment or contact with a patient
Guide tube, projectile and O-rings
replace
after 1,000,000 shocks (overhaul the handpiece)
Caution!
Flammable and volatile cleaning and disinfecting agents may generate an explosive atmosphere.
Disconnect the handpiece from the control unit before starting any cleaning or maintenance work.
Attention!
It is essential that no fluid be permitted to penetrate either the device or its tubing.
*Falcon is a registered trade mark of Storz medical AG
Summary of Contents for Intelect RPW 2
Page 32: ...32 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Select transmitter ...
Page 50: ...50 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Delete one session ...
Page 55: ...55 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Delete one protocol ...
Page 61: ...61 USER INTERFACE EN INTELECT RPW 2 USER MANUAL 4 Full screen image Close full screen mode ...