GENERAL WARNINGS AND PRECAUTIONS
EN
INTELECT
®
RPW 2 USER MANUAL
• This device should be operated at 10°C to +40°C and 30% to 75% Relative Humidity (no condensation). The device should
be transported and stored at -20°C to +60°C and 10% to 90% Relative Humidity (no condensation).
• Atmospheric pressure requirements: 70 kPa to 106 kPa.
• The Falcon handpiece should be operated at 10°C to +40°C and 5% to 95% Relative Humidity (no condensation).
The handpiece should be transported and stored at 0°C to +60°C and 5% to 95% Relative Humidity (no condensation).
• The V-Actor HF handpiece should be operated at 10°C to +30°C and 5% to 95% Relative Humidity (no condensation). The
handpiece should be transported and stored at 0°C to +60°C and 5% to 55% Relative Humidity (no condensation).
• The time required for Intelect RPW 2 to warm from the minimum storage temperature between uses until Intelect RPW
2 is ready for its intended use when the ambient temperature is 20°C is 5 hours.
• The time required for Intelect RPW 2 to cool from the maximum storage temperature between uses until Intelect RPW 2
is ready for its Intended use when the ambient temperature is 20 °C is 5 hours
• Under single fault condition and extreme usage conditions, the maximum temperature of the hand piece can reach 47°C.
There is no particular health risk associated with this temperature besides your comfort.
• Use of parts or materials other than DJO’s can degrade minimum safety.
• Connect to this device only items and equipment that have been specified in this IFU as part of the Intelect RPW 2
System or that have been specified as being compatible with the Intelect RPW 2 System.
• DO NOT disassemble, modify, or remodel the device or accessories. This may cause device damage, malfunction, electrical
shock, fire, or personal injury.
• Before using the therapy device, the operator must ascertain that it is in correct working order and operating condition.
The cables and connectors, in particular, must be checked for signs of damage. The device cannot be used if any part is
damaged as there is a risk of electrical shock.
• Before each use always offer the patient hearing protection.
• Apply a sufficient amount of coupling gel to the patient’s skin in the treatment area and to the shock transmitter.
• The maximum energy level used during treatment must not cause the patient undue pain under any circumstances.
• DO NOT apply more than 300 shocks to the same spot during treatment.
• The surface of the transmitter can reach 47°C. Extended skin contact can cause minor burns. Interrupt therapy after a
maximum of 6000 pulses to allow the transmitter to cool.
• Operating the device at pressures higher than 3 bar without an impact surface can result in damage to the handset.
• To avoid a trip hazard, keep all accessories and their cords separated during treatment by using the hook located on the
side of the device.
• Disconnect the instrument from the mains before starting any cleaning or overhaul work.
• Danger of injury due to pulse triggering when handpiece is open.
• Disconnect the handpiece from the control device before starting any cleaning or maintenance work.
• There are no user-serviceable parts inside the device. If a malfunction occurs, discontinue use immediately and contact
your local DJO office via www.djoglobal.com for assistance and service information.
• Do not use while sleeping or operating heavy equipment.
CAUTION
Summary of Contents for Intelect RPW 2
Page 32: ...32 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Select transmitter ...
Page 50: ...50 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Delete one session ...
Page 55: ...55 USER INTERFACE EN INTELECT RPW 2 USER MANUAL Delete one protocol ...
Page 61: ...61 USER INTERFACE EN INTELECT RPW 2 USER MANUAL 4 Full screen image Close full screen mode ...